{"id":2658,"date":"2026-08-13T00:08:37","date_gmt":"2026-08-13T00:08:37","guid":{"rendered":"https:\/\/www.pharmacopoeiatest.com\/?p=2658"},"modified":"2026-08-13T00:08:40","modified_gmt":"2026-08-13T00:08:40","slug":"tubing-stiffness-tester-for-iso-9626","status":"publish","type":"post","link":"https:\/\/www.pharmacopoeiatest.com\/lv\/tubing-stiffness-tester-for-iso-9626\/","title":{"rendered":"Cauru\u013cu stingr\u012bbas test\u0113t\u0101js: ISO 9626 medic\u012bnisko cauru\u013cu stingr\u012bbas tests"},"content":{"rendered":"<h2 class=\"wp-block-heading\">Kas ir cauru\u013cu stingr\u012bbas test\u0113t\u0101js?<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">A <strong>cauru\u013cu stingr\u012bbas test\u0113t\u0101js<\/strong> nov\u0113rt\u0113 medic\u012bnisko cauru\u013cu iztur\u012bbu pret deform\u0101ciju, uz t\u0101m iedarbojoties ar noteiktu sp\u0113ku. Attiec\u012bb\u0101 uz injekciju adatu caurul\u0113m un cit\u0101m medic\u012bniskaj\u0101m caurul\u0113m stingr\u012bba tie\u0161i ietekm\u0113 elast\u012bbu, ievad\u012b\u0161anas \u012bpa\u0161\u012bbas, meh\u0101nisko uzticam\u012bbu un dro\u0161\u012bbu. \u0160is <strong>TS-01 cauru\u013cu stingr\u012bbas test\u0113t\u0101js<\/strong> uz atbalst\u012bta cauru\u013cveida parauga centru iedarbojas ar kontrol\u0113tu sp\u0113ku, kas v\u0113rsts uz leju, un m\u0113ra t\u0101 rezult\u0101t\u0101 radu\u0161os deform\u0101ciju. \u0160\u012b test\u0113\u0161anas metode nodro\u0161ina <strong>medic\u012bnisko cauru\u013cu stingr\u012bbas p\u0101rbaude<\/strong> saska\u0146\u0101 ar <strong>ISO 9626<\/strong> un <strong>GB\/T 18457<\/strong>.<\/p>\n\n\n\n<div class=\"wp-block-uagb-image aligncenter uagb-block-f73b64ff wp-block-uagb-image--layout-default wp-block-uagb-image--effect-static wp-block-uagb-image--align-center\"><figure class=\"wp-block-uagb-image__figure\"><img loading=\"lazy\" decoding=\"async\" srcset=\"\" sizes=\"auto, (max-width: 480px) 150px\" src=\"https:\/\/www.pharmacopoeiatest.com\/wp-content\/uploads\/2026\/08\/Tubing-Stiffness-Tester: ISO-9626-Medical-Tubing-Stiffness-Test.webp\" alt=\"Cauru\u013cu stingr\u012bbas test\u0113t\u0101js: ISO 9626 medic\u012bnisko cauru\u013cu stingr\u012bbas tests\" class=\"uag-image-2661\" width=\"428\" height=\"428\" title=\"Cauru\u013cu stingr\u012bbas test\u0113t\u0101js: ISO 9626 medic\u012bnisko cauru\u013cu stingr\u012bbas tests\" role=\"img\" data-srcset=\"\"\/><\/figure><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">K\u0101p\u0113c cauru\u013cu stingr\u012bba ir svar\u012bga medic\u012bnas ier\u012bc\u0113m?<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Medic\u012bniskaj\u0101m caurul\u012bt\u0113m ir j\u0101nodro\u0161ina atbilsto\u0161s l\u012bdzsvars starp <strong>iztur\u012bba un elast\u012bba<\/strong>. Ja caurule ir p\u0101r\u0101k elast\u012bga, t\u0101 var saliekties, deform\u0113ties vai tikt boj\u0101ta t\u0101s p\u0101rvieto\u0161anas un lieto\u0161anas laik\u0101. Ja t\u0101 ir p\u0101r\u0101k st\u012bva, tai var neb\u016bt pietiekamas elast\u012bbas, un noteiktos apst\u0101k\u013cos t\u0101 var k\u013c\u016bt vair\u0101k pak\u013cauta trauslas l\u016bzuma riskam.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Medic\u012bnisko ier\u012b\u010du ra\u017eot\u0101jiem cauru\u013cu stingr\u012bbas test\u0113\u0161ana var pal\u012bdz\u0113t nov\u0113rt\u0113t:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Iztur\u012bba pret deform\u0101ciju slodzes ietekm\u0113<\/li>\n\n\n\n<li>Elast\u012bba un meh\u0101nisk\u0101s \u012bpa\u0161\u012bbas<\/li>\n\n\n\n<li>Ra\u017eo\u0161anas partiju savstarp\u0113j\u0101 atbilst\u012bba<\/li>\n\n\n\n<li>Cauru\u013cu piem\u0113rot\u012bba medic\u012bnisk\u0101m vajadz\u012bb\u0101m<\/li>\n\n\n\n<li>Atbilst\u012bba piem\u0113rojamaj\u0101m specifik\u0101cij\u0101m un standartiem<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">T\u0101p\u0113c kontrol\u0113ta <strong>medic\u012bnisko ier\u012b\u010du cauru\u013cu stingr\u012bbas p\u0101rbaude<\/strong> sniedz noder\u012bgus meh\u0101niskos datus materi\u0101lu izv\u0113lei, procesa kontrolei, izejvielu p\u0101rbaudei un gal\u012bgajai kvalit\u0101tes p\u0101rbaudei.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">K\u0101 darbojas cauru\u013cu stingr\u012bbas test\u0113t\u0101js?<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Pamatprincips ir vienk\u0101r\u0161s. Noteikta garuma caurule ir <strong>nostiprin\u0101ts abos galos<\/strong>, vienlaikus uz t\u0101 centru tiek pielikts kontrol\u0113ts sp\u0113ks. P\u0113c tam instruments nosaka, cik t\u0101lu caurule izliecas.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Lai veiktu \u0161o m\u0113r\u012bjumu, TS-01 izmanto slodzes dev\u0113ju un speci\u0101lu virzuli. Virzulis uz cauruli iedarbojas ar sp\u0113ku vertik\u0101l\u0101 virzien\u0101. T\u0101 apak\u0161\u0113j\u0101 gal\u0101 ir ieb\u016bv\u0113ts <strong>koniska konfigur\u0101cija ar virsm\u0101m, kas viena pret otru sl\u012bpst 60\u00b0 le\u0146\u0137\u012b<\/strong>, kop\u0101 ar cilindrisku virsmu, kuras r\u0101diuss ir 1 mm un garums vismaz 5 mm.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Sist\u0113ma var pielietot sp\u0113ku l\u012bdz pat <strong>60 N<\/strong> ar sp\u0113ka precizit\u0101ti <strong>\u00b10,01 N<\/strong>, savuk\u0101rt caurules lieci var izm\u0113r\u012bt l\u012bdz <strong>0,01 mm<\/strong>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">\u0160o pamatattiec\u012bbu konceptu\u0101li var izteikt \u0161\u0101di:<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Pieliktais sp\u0113ks \u2192 Caurules liece \u2192 St\u012bvuma nov\u0113rt\u0113\u0161ana<\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Maz\u0101ks novirziens pie t\u0101 pa\u0161a pielikt\u0101 sp\u0113ka parasti liecina par liel\u0101ku iztur\u012bbu pret deform\u0101ciju, savuk\u0101rt liel\u0101ks novirziens nor\u0101da uz elast\u012bg\u0101ku cauru\u013cveida paraugu.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">ISO 9626 cauru\u013cu stingr\u012bbas p\u0101rbaude<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>ISO 9626<\/strong> ir svar\u012bgs atsauces dokuments attiec\u012bb\u0101 uz meh\u0101niskaj\u0101m pras\u012bb\u0101m un test\u0113\u0161anu ner\u016bs\u0113jo\u0161\u0101 t\u0113rauda adatu caurul\u0113m, ko izmanto medic\u012bnisko ier\u012b\u010du ra\u017eo\u0161an\u0101. Tas nodro\u0161ina standartiz\u0113tu sist\u0113mu cauru\u013cu \u012bpa\u0161\u012bbu nov\u0113rt\u0113\u0161anai, tostarp meh\u0101niskaj\u0101m \u012bpa\u0161\u012bb\u0101m, kas ir b\u016btiskas to paredz\u0113tajai lieto\u0161anai.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Veicot <strong>ISO 9626 cauru\u013cu stingr\u012bbas p\u0101rbaude<\/strong>, ir \u013coti svar\u012bgi kontrol\u0113t p\u0101rbaudes sp\u0113ku, parauga atbalstu, slodzes uzlik\u0161anas poz\u012bciju un deform\u0101cijas m\u0113r\u012bjumus. Izmantojot speci\u0101lu <strong>cauru\u013cu stingr\u012bbas test\u0113t\u0101js<\/strong> pal\u012bdz laboratorij\u0101m nodro\u0161in\u0101t nemain\u012bgus test\u0113\u0161anas apst\u0101k\u013cus un samazin\u0101t no operatora atkar\u012bg\u0101s novirzes.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>GB\/T 18457<\/strong> ir v\u0113l viens piem\u0113rojams standarts ner\u016bs\u0113jo\u0161\u0101 t\u0113rauda adatu caurul\u0113m, un to var izmantot k\u0101 atsauci attiec\u012bgaj\u0101m test\u0113\u0161anas pras\u012bb\u0101m.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><thead><tr><th class=\"has-text-align-center\" data-align=\"center\">Prece<\/th><th class=\"has-text-align-center\" data-align=\"center\">TS-01 funkcijas<\/th><\/tr><\/thead><tbody><tr><td class=\"has-text-align-center\" data-align=\"center\">Piem\u0113rojamie standarti<\/td><td class=\"has-text-align-center\" data-align=\"center\">ISO 9626, GB\/T 18457<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">Maksim\u0101lais fiks\u0113tais sp\u0113ks<\/td><td class=\"has-text-align-center\" data-align=\"center\">60 N vai p\u0113c nepiecie\u0161am\u012bbas<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">Sp\u0113ka precizit\u0101te<\/td><td class=\"has-text-align-center\" data-align=\"center\">\u00b10,01 N<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">Novirzes precizit\u0101te<\/td><td class=\"has-text-align-center\" data-align=\"center\">0,01 mm<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">Testa \u0101trums<\/td><td class=\"has-text-align-center\" data-align=\"center\">1\u2013500 mm\/min<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">Kontrole<\/td><td class=\"has-text-align-center\" data-align=\"center\">PLC<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">Darb\u012bba<\/td><td class=\"has-text-align-center\" data-align=\"center\">7 collu HMI sk\u0101rienekr\u0101ns<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">Ener\u0123ijas padeve<\/td><td class=\"has-text-align-center\" data-align=\"center\">110\u2013220 V<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">K\u0101 tiek veikts cauru\u013cu stingr\u012bbas tests?<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Tipiska testa sec\u012bba ietver \u0161\u0101dus posmus:<\/p>\n\n\n\n<ol class=\"wp-block-list\">\n<li><strong>Sagatavojiet paraugu<\/strong> saska\u0146\u0101 ar piem\u0113rojamaj\u0101m test\u0113\u0161anas pras\u012bb\u0101m.<\/li>\n\n\n\n<li><strong>Novietojiet cauruli<\/strong> uz atbalsta ier\u012bces ar nor\u0101d\u012bto p\u0101rbaudes garumu.<\/li>\n\n\n\n<li><strong>Izvietojiet virzuli<\/strong> ar caurules centru.<\/li>\n\n\n\n<li><strong>Iestatiet testa parametrus<\/strong>, ieskaitot sp\u0113ku vai testa \u0101trumu.<\/li>\n\n\n\n<li><strong>Piem\u0113rojiet sp\u0113ku uz leju<\/strong> caur slodzes dev\u0113ju un virzuli.<\/li>\n\n\n\n<li><strong>Izm\u0113r\u012bt cauru\u013cvada izlieci<\/strong> iekrau\u0161anas laik\u0101.<\/li>\n\n\n\n<li><strong>Re\u0123istr\u0113jiet testa rezult\u0101tu<\/strong> kvalit\u0101tes nov\u0113rt\u0113\u0161anai un sal\u012bdzin\u0101\u0161anai.<\/li>\n<\/ol>\n\n\n\n<p class=\"wp-block-paragraph\">TS-01 p\u0113c test\u0113\u0161anas autom\u0101tiski atgrie\u017eas s\u0101kotn\u0113j\u0101 st\u0101vokl\u012b, kas var uzlabot test\u0113\u0161anas efektivit\u0101ti gad\u012bjumos, kad laboratorij\u0101m ir j\u0101izv\u0113rt\u0113 vair\u0101ki paraugi.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">K\u0101du apr\u012bkojumu izmanto medic\u012bnisko cauru\u013cu stingr\u012bbas p\u0101rbaud\u0113m?<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Speci\u0101li paredz\u0113ts <strong>cauru\u013cu stingr\u012bbas test\u0113t\u0101js<\/strong> parasti sast\u0101v no kontrol\u0113ta testa r\u0101mja, slodzes m\u0113r\u012b\u0161anas sist\u0113mas, virzuli, parauga tur\u0113t\u0101ja un programmat\u016bras vai saskarnes testa apst\u0101k\u013cu iestat\u012b\u0161anai un re\u0123istr\u0113\u0161anai.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">TS-01 ietver:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li><strong>PLC kontrol\u0113ta darb\u012bba<\/strong><\/li>\n\n\n\n<li><strong>7 collu HMI sk\u0101rienekr\u0101ns<\/strong><\/li>\n\n\n\n<li><strong>Precizit\u0101tes lod\u012b\u0161u svina skr\u016bve<\/strong><\/li>\n\n\n\n<li><strong>Main\u012bgs testa \u0101trums<\/strong><\/li>\n\n\n\n<li><strong>Aizsardz\u012bba pret p\u0101rslodzi<\/strong><\/li>\n\n\n\n<li><strong>Autom\u0101tisk\u0101s atgrie\u0161an\u0101s funkcija<\/strong><\/li>\n\n\n\n<li><strong>Mikroprinteris<\/strong><\/li>\n\n\n\n<li>Vair\u0101ki stiprin\u0101jumi sp\u0113ka iedarb\u012bbas uz leju p\u0101rbaud\u0113m<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">\u0160\u012bs funkcijas \u013cauj laboratorij\u0101m nodro\u0161in\u0101t atk\u0101rtojamus test\u0113\u0161anas apst\u0101k\u013cus, vienlaikus vienk\u0101r\u0161ojot ikdienas darb\u012bbu.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">K\u0101p\u0113c izmantot specializ\u0113tu cauru\u013cu stingr\u012bbas testeri?<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Veicot deform\u0101cijas m\u0113r\u012bjumus manu\u0101li, var rasties novirzes slodzes sp\u0113k\u0101, parauga novietojum\u0101 un m\u0113r\u012bjumu precizit\u0101t\u0113. Speci\u0101li \u0161im nol\u016bkam paredz\u0113ts instruments nodro\u0161ina kontrol\u0113tu meh\u0101nisko slodzi un digit\u0101lus m\u0113r\u012bjumus, t\u0101d\u0113j\u0101di rezult\u0101tus ir viegl\u0101k atk\u0101rtot un sal\u012bdzin\u0101t.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Laboratorij\u0101m, kas str\u0101d\u0101 ar <strong>medic\u012bnisko cauru\u013cu stingr\u012bbas p\u0101rbaude<\/strong>, galven\u0101s priek\u0161roc\u012bbas ir \u0161\u0101das:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li><strong>Kontrol\u0113ta sp\u0113ka pielieto\u0161ana<\/strong><\/li>\n\n\n\n<li><strong>Prec\u012bzs novirzes m\u0113r\u012bjums<\/strong><\/li>\n\n\n\n<li><strong>Atk\u0101rtojami test\u0113\u0161anas apst\u0101k\u013ci<\/strong><\/li>\n\n\n\n<li><strong>Regul\u0113jams testa \u0101trums<\/strong><\/li>\n\n\n\n<li><strong>Uz standartiem balst\u012bta test\u0113\u0161ana<\/strong><\/li>\n\n\n\n<li><strong>Efekt\u012bva ikdienas kvalit\u0101tes kontrole<\/strong><\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">TS-01 atbalsta gan standarta test\u0113\u0161anas pras\u012bbas, gan individu\u0101las test\u0113\u0161anas vajadz\u012bbas, nodro\u0161inot maksim\u0101lo sp\u0113ku 60 N un konfigur\u0113jamus test\u0113\u0161anas apst\u0101k\u013cus.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Secin\u0101jums<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Cauru\u013cu stingr\u012bba ir svar\u012bga medic\u012bnisko cauru\u013cu meh\u0101nisk\u0101 \u012bpa\u0161\u012bba, jo t\u0101 ietekm\u0113 elast\u012bbu, iztur\u012bbu pret deform\u0101cij\u0101m un funkcion\u0101lo uzticam\u012bbu. A <strong>cauru\u013cu stingr\u012bbas test\u0113t\u0101js<\/strong> nodro\u0161ina kontrol\u0113tu metodi, k\u0101 pielietot sp\u0113ku uz balst\u012btu cauru\u013cveida paraugu un izm\u0113r\u012bt t\u0101 deform\u0101ciju. Ar atbalstu <strong><a href=\"https:\/\/www.iso.org\/standard\/60480.html\" target=\"_blank\" rel=\"noreferrer noopener\">ISO 9626<\/a><\/strong> un <strong>GB\/T 18457<\/strong>, TS-01 ir paredz\u0113ts laboratorij\u0101m un ra\u017eot\u0101jiem, kas veic medic\u012bnisko cauru\u013cu stingr\u012bbas nov\u0113rt\u0113\u0161anu, kvalit\u0101tes kontroli, materi\u0101lu izstr\u0101di un ra\u017eo\u0161anas p\u0101rbaudi.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Port\u0101ls <strong>TS-01 cauru\u013cu stingr\u012bbas test\u0113t\u0101js<\/strong> nodro\u0161ina kontrol\u0113tu sp\u0113ka pielik\u0161anu, prec\u012bzu deform\u0101cijas m\u0113r\u012b\u0161anu un elast\u012bgus testa iestat\u012bjumus ikdienas laboratorijas test\u0113\u0161anai. Ja j\u016bsu darb\u012bbas jom\u0101 ietilpst ISO 9626, GB\/T 18457, medic\u012bnisko cauru\u013cu stingr\u012bbas test\u0113\u0161ana vai specializ\u0113ta cauru\u013cu nov\u0113rt\u0113\u0161ana, <strong>\u0160\u016bnu instrumenti <\/strong>var nodro\u0161in\u0101t atbilsto\u0161as test\u0113\u0161anas iek\u0101rtas, stiprin\u0101jumus un individu\u0101li piel\u0101gotus test\u0113\u0161anas risin\u0101jumus.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>","protected":false},"excerpt":{"rendered":"<p>Kas ir cauru\u013cu stingr\u012bbas test\u0113t\u0101js? Cauru\u013cu stingr\u012bbas test\u0113t\u0101js m\u0113ra medic\u012bnisko cauru\u013cu pretest\u012bbu pret deform\u0101ciju, uz t\u0101m iedarbojoties ar noteiktu sp\u0113ku. Attiec\u012bb\u0101 uz injekciju adatu caurul\u0113m un cit\u0101m medic\u012bniskaj\u0101m caurul\u0113m stingr\u012bba tie\u0161i ietekm\u0113 elast\u012bbu, ievad\u012b\u0161anas \u012bpa\u0161\u012bbas, meh\u0101nisko uzticam\u012bbu un dro\u0161\u012bbu. Cauru\u013cu stingr\u012bbas test\u0113t\u0101js TS-01 piem\u0113ro kontrol\u0113tu sp\u0113ku uz leju caurules centram [\u2026]<\/p>","protected":false},"author":5,"featured_media":2661,"comment_status":"closed","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_uag_custom_page_level_css":"","site-sidebar-layout":"default","site-content-layout":"","ast-site-content-layout":"default","site-content-style":"default","site-sidebar-style":"default","ast-global-header-display":"","ast-banner-title-visibility":"","ast-main-header-display":"","ast-hfb-above-header-display":"","ast-hfb-below-header-display":"","ast-hfb-mobile-header-display":"","site-post-title":"","ast-breadcrumbs-content":"","ast-featured-img":"","footer-sml-layout":"","theme-transparent-header-meta":"","adv-header-id-meta":"","stick-header-meta":"","header-above-stick-meta":"","header-main-stick-meta":"","header-below-stick-meta":"","astra-migrate-meta-layouts":"set","ast-page-background-enabled":"default","ast-page-background-meta":{"desktop":{"background-color":"var(--ast-global-color-4)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""},"tablet":{"background-color":"","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""},"mobile":{"background-color":"","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""}},"ast-content-background-meta":{"desktop":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""},"tablet":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""},"mobile":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""}},"footnotes":""},"categories":[26],"tags":[],"class_list":["post-2658","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-blog"],"acf":[],"uagb_featured_image_src":{"full":["https:\/\/www.pharmacopoeiatest.com\/wp-content\/uploads\/2026\/08\/Tubing-Stiffness-Tester\uff1aISO-9626-Medical-Tubing-Stiffness-Test.webp",600,600,false],"thumbnail":["https:\/\/www.pharmacopoeiatest.com\/wp-content\/uploads\/2026\/08\/Tubing-Stiffness-Tester\uff1aISO-9626-Medical-Tubing-Stiffness-Test-150x150.webp",150,150,true],"medium":["https:\/\/www.pharmacopoeiatest.com\/wp-content\/uploads\/2026\/08\/Tubing-Stiffness-Tester\uff1aISO-9626-Medical-Tubing-Stiffness-Test-300x300.webp",300,300,true],"medium_large":["https:\/\/www.pharmacopoeiatest.com\/wp-content\/uploads\/2026\/08\/Tubing-Stiffness-Tester\uff1aISO-9626-Medical-Tubing-Stiffness-Test.webp",600,600,false],"large":["https:\/\/www.pharmacopoeiatest.com\/wp-content\/uploads\/2026\/08\/Tubing-Stiffness-Tester\uff1aISO-9626-Medical-Tubing-Stiffness-Test.webp",600,600,false],"1536x1536":["https:\/\/www.pharmacopoeiatest.com\/wp-content\/uploads\/2026\/08\/Tubing-Stiffness-Tester\uff1aISO-9626-Medical-Tubing-Stiffness-Test.webp",600,600,false],"2048x2048":["https:\/\/www.pharmacopoeiatest.com\/wp-content\/uploads\/2026\/08\/Tubing-Stiffness-Tester\uff1aISO-9626-Medical-Tubing-Stiffness-Test.webp",600,600,false],"trp-custom-language-flag":["https:\/\/www.pharmacopoeiatest.com\/wp-content\/uploads\/2026\/08\/Tubing-Stiffness-Tester\uff1aISO-9626-Medical-Tubing-Stiffness-Test-12x12.webp",12,12,true]},"uagb_author_info":{"display_name":"Editor","author_link":"https:\/\/www.pharmacopoeiatest.com\/lv\/author\/editor\/"},"uagb_comment_info":0,"uagb_excerpt":"What Is a Tubing Stiffness Tester? A tubing stiffness tester measures the resistance of medical tubing to deformation under a specified force. For hypodermic needle tubing and other medical tubing, stiffness directly affects flexibility, insertion performance, mechanical reliability, and safety. The TS-01 Tubing Stiffness Tester applies a controlled downward force to the center of a&hellip;","_links":{"self":[{"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/posts\/2658","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/users\/5"}],"replies":[{"embeddable":true,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/comments?post=2658"}],"version-history":[{"count":5,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/posts\/2658\/revisions"}],"predecessor-version":[{"id":2664,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/posts\/2658\/revisions\/2664"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/media\/2661"}],"wp:attachment":[{"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/media?parent=2658"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/categories?post=2658"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/tags?post=2658"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}