{"id":2570,"date":"2026-03-03T08:38:48","date_gmt":"2026-03-03T08:38:48","guid":{"rendered":"https:\/\/www.pharmacopoeiatest.com\/?p=2570"},"modified":"2026-03-03T08:38:50","modified_gmt":"2026-03-03T08:38:50","slug":"needle-toughness-test","status":"publish","type":"post","link":"https:\/\/www.pharmacopoeiatest.com\/lv\/needle-toughness-test\/","title":{"rendered":"Adatu iztur\u012bbas tests: stingr\u012bbas un l\u016bzuma iztur\u012bbas p\u0101rbaude saska\u0146\u0101 ar standartu ISO 9626"},"content":{"rendered":"<h2 class=\"wp-block-heading\">Ievads adatas iztur\u012bbas p\u0101rbaud\u0113<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Port\u0101ls <strong>Adatas iztur\u012bbas tests<\/strong> ir b\u016btisks, lai nov\u0113rt\u0113tu adatu cauru\u013cu meh\u0101nisk\u0101s \u012bpa\u0161\u012bbas, kuras izmanto hipoderm\u0101lo adatu un citu medic\u012bnisko ier\u012b\u010du ra\u017eo\u0161an\u0101. Viens no \u0161\u012b testa galvenajiem aspektiem ir p\u0101rbaud\u012bt <strong>cauru\u013cu iztur\u012bba pret l\u016b\u0161anu<\/strong>, kas ir \u0101rk\u0101rt\u012bgi svar\u012bgi, lai nodro\u0161in\u0101tu medic\u012bnisko ier\u012b\u010du dro\u0161\u012bbu un darb\u012bbas efektivit\u0101ti. \u0160is tests, kas ietver gan <strong>st\u012bvums<\/strong> un <strong>iztur\u012bba pret l\u016bzumiem<\/strong>, tiek veikta saska\u0146\u0101 ar atz\u012btiem standartiem, piem\u0113ram, <strong>ISO 9626<\/strong>.<\/p>\n\n\n\n<div class=\"wp-block-uagb-image uagb-block-fe099ac4 wp-block-uagb-image--layout-default wp-block-uagb-image--effect-static wp-block-uagb-image--align-none\"><figure class=\"wp-block-uagb-image__figure\"><a class=\"\" href=\"https:\/\/www.syringetest.com\/standards\/iso-9626\" target=\"_blank\" rel=\"noreferrer noopener\"><img loading=\"lazy\" decoding=\"async\"  sizes=\"auto, (max-width: 480px) 150px\" src=\"https:\/\/www.pharmacopoeiatest.com\/wp-content\/uploads\/2026\/03\/Needle-Toughness-Test-Stiffness-and-Breakage-Resistance-Testing-with-ISO-9626.webp\" alt=\"\" width=\"885\" height=\"885\" title=\"\" role=\"img\"\/><\/a><\/figure><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">ISO 9626 standarta izpratne un t\u0101 noz\u012bme<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>ISO 9626:2016<\/strong> nosaka pras\u012bbas un p\u0101rbaudes metodes cietaj\u0101m ner\u016bs\u0113jo\u0161\u0101 t\u0113rauda adatu caurul\u012bt\u0113m. \u0160\u012bs caurul\u012btes galvenok\u0101rt tiek izmantotas medic\u012bniskaj\u0101s ier\u012bc\u0113s, tostarp injekciju adat\u0101s, un t\u0101s ir b\u016btiskas lieto\u0161anai cilv\u0113kiem. \u0160is <strong>ISO 9626<\/strong> Standarts nosaka cauru\u013cvadiem nepiecie\u0161amos izm\u0113rus, meh\u0101nisk\u0101s \u012bpa\u0161\u012bbas un ekspluat\u0101cijas p\u0101rbaudes.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Port\u0101ls <strong>ISO 9626<\/strong> Standarts attiecas uz ner\u016bs\u0113jo\u0161\u0101 t\u0113rauda caurul\u0113m, kuru diametrs ir no 0,18 mm (34 Gauge) l\u012bdz 3,4 mm (10 Gauge). Taj\u0101 ir noteiktas p\u0101rbaudes vadl\u012bnijas <strong>st\u012bvums<\/strong>, <strong>iztur\u012bba pret l\u016b\u0161anu<\/strong>, k\u0101 ar\u012b citas meh\u0101nisk\u0101s \u012bpa\u0161\u012bbas, lai nodro\u0161in\u0101tu, ka caurule sp\u0113j iztur\u0113t spriedzi un deform\u0101cijas, kas var rasties medic\u012bnisko proced\u016bru laik\u0101.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Adatu cauru\u013cu iztur\u012bbas pret l\u016bzumiem noz\u012bme<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Viena no galvenaj\u0101m probl\u0113m\u0101m medic\u012bnisko ier\u012b\u010du ra\u017eo\u0161an\u0101 ir nodro\u0161in\u0101t, lai caurules b\u016btu iztur\u012bgas pret l\u016bzumiem. A <strong>adatas caurules iztur\u012bbas pret l\u016b\u0161anu p\u0101rbaude<\/strong> nov\u0113rt\u0113 adatas caurules sp\u0113ju iztur\u0113t meh\u0101nisko slodzi, nepl\u012bstot. Test\u0101 uz cauruli tiek pielikts lieces sp\u0113ks, imit\u0113jot meh\u0101nisk\u0101s slodzes, k\u0101das caurule var\u0113tu piedz\u012bvot re\u0101los lieto\u0161anas apst\u0101k\u013cos.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">\u0160aj\u0101 test\u0101 caurule tiek pak\u013cauta noteiktam lieces ciklu skaitam, katru reizi liecot cauruli vien\u0101 virzien\u0101 un p\u0113c tam mainot lieces virzienu. P\u0113c testa ciklu pabeig\u0161anas caurule tiek p\u0101rbaud\u012bta, lai konstat\u0113tu redzamus boj\u0101jumus. <strong>cauru\u013cu iztur\u012bba pret l\u016b\u0161anu<\/strong> ir \u0101rk\u0101rt\u012bgi svar\u012bgi, jo pat nelieli defekti vai l\u016bzumi var apdraud\u0113t medic\u012bnisko ier\u012b\u010du dro\u0161\u012bbu un funkcionalit\u0101ti.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Cauru\u013cu stingr\u012bba: b\u016btisks parametrs medic\u012bnisko ier\u012b\u010du ra\u017eo\u0161an\u0101<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Port\u0101ls <strong>cauru\u013cvadu stingr\u012bba<\/strong> ir v\u0113l viena b\u016btiska meh\u0101nisk\u0101 \u012bpa\u0161\u012bba, kas ietekm\u0113 medic\u012bnisko ier\u012b\u010du darb\u012bbu, jo \u012bpa\u0161i t\u0101d\u0101s lietojuma jom\u0101s k\u0101 injekciju adatas. <strong>St\u012bvums<\/strong> attiecas uz caurules iztur\u012bbu pret deform\u0101ciju, kad uz to tiek pielikts sp\u0113ks. A <strong>cauru\u013cu st\u012bvuma tests<\/strong> to veic, pieliekot noteiktu sp\u0113ku uz leju caurules centram, kam\u0113r t\u0101 ir atbalst\u012bta abos galos.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Saska\u0146\u0101 ar <strong>ISO 9626<\/strong>, izmantojot virzuli, tiek pielikts sp\u0113ks caurules centr\u0101, un tiek m\u0113r\u012bta caurules deform\u0101cija. Deform\u0101cijas pak\u0101pe liecina par to, cik stingra vai elast\u012bga ir caurule. Medic\u012bniskaj\u0101m ier\u012bc\u0113m pareizais stingr\u012bbas l\u012bmenis ir b\u016btisks, lai nodro\u0161in\u0101tu, ka adata darbojas uzticami, neradot risku, ka t\u0101 lieto\u0161anas laik\u0101 salieksies vai sal\u016bz\u012bs.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Galven\u0101s adatu iztur\u012bbas p\u0101rbaudes metodes<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">1. Cauru\u013cu stingr\u012bbas p\u0101rbaude<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">... <strong>cauru\u013cu st\u012bvuma tests<\/strong>, uz caurules centru tiek pielikts noteikts sp\u0113ks, un tiek izm\u0113r\u012bta novirze \u0161aj\u0101 punkt\u0101. \u0160is tests simul\u0113 to, k\u0101 adata darbosies, kad medic\u012bnisko proced\u016bru laik\u0101 tiks pielikts spiediens. \u0160is <strong>st\u012bvuma p\u0101rbaudes apar\u0101ts<\/strong> ir paredz\u0113ts, lai pieliktu sp\u0113ku l\u012bdz 60 N un izm\u0113r\u012btu novirzi ar precizit\u0101ti \u00b10,01 mm. \u0160is tests nodro\u0161ina, ka adata nav ne p\u0101r\u0101k st\u012bva, kas var\u0113tu ietekm\u0113t t\u0101s lieto\u0161anas \u0113rtumu, ne ar\u012b p\u0101r\u0101k elast\u012bga, kas var\u0113tu apdraud\u0113t t\u0101s integrit\u0101ti.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Testa proced\u016bra noris \u0161\u0101di:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Caurule tiek novietota uz p\u0101rbaudes iek\u0101rtas, kas uz caurules centru iedarbojas ar sp\u0113ku.<\/li>\n\n\n\n<li>Sp\u0113ks tiek pielikts ar \u0101trumu no 1 mm\/min l\u012bdz 10 mm\/min.<\/li>\n\n\n\n<li>Lieciens tiek izm\u0113r\u012bts un re\u0123istr\u0113ts ar precizit\u0101ti l\u012bdz 0,01 mm.<\/li>\n<\/ul>\n\n\n\n<h3 class=\"wp-block-heading\">2. Cauru\u013cu iztur\u012bbas pret l\u016bzumiem p\u0101rbaude<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Port\u0101ls <strong>cauru\u013cu iztur\u012bba pret l\u016b\u0161anu<\/strong> Test\u0101 caurules tiek saliektas noteiktos le\u0146\u0137os, imit\u0113jot re\u0101los spriegumus, k\u0101diem caurules var tikt pak\u013cautas. Test\u0101 ietilpst:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Caurules viena gala nostiprin\u0101\u0161ana un sp\u0113ka pielik\u0161ana standartos nor\u0101d\u012btaj\u0101 punkt\u0101.<\/li>\n\n\n\n<li>Caurules liek\u0161ana noteikt\u0101 le\u0146\u0137\u012b, parasti 25\u00b0 standarta sienu biezuma caurul\u0113m.<\/li>\n\n\n\n<li>Sp\u0113ka virziena mai\u0146a, lai cauruli saliektu pret\u0113j\u0101 virzien\u0101.<\/li>\n\n\n\n<li>\u0160o procesu atk\u0101rto 20 ciklu garum\u0101, lai p\u0101rbaud\u012btu cauru\u013cvada iztur\u012bbu.<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">P\u0113c p\u0101rbaudes caurules tiek vizu\u0101li p\u0101rbaud\u012btas, lai konstat\u0113tu to l\u016bzuma vai boj\u0101jumu paz\u012bmes.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">K\u0101p\u0113c ir svar\u012bgi veikt test\u0113\u0161anu saska\u0146\u0101 ar ISO 9626<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Test\u0113\u0161ana <strong>adatas iztur\u012bba<\/strong> saska\u0146\u0101 ar <strong>ISO 9626<\/strong> nodro\u0161ina, ka medic\u012bniskaj\u0101s ier\u012bc\u0113s izmantot\u0101s caurules atbilst starptautiskajiem dro\u0161\u012bbas, veiktsp\u0113jas un iztur\u012bbas standartiem. \u0160ie testi ir b\u016btiski ra\u017eot\u0101jiem, kuri v\u0113las garant\u0113t, ka to izstr\u0101d\u0101jumi lieto\u0161anas laik\u0101 nedarbosies nepareizi, kas var\u0113tu izrais\u012bt traumas vai negat\u012bvi ietekm\u0113t \u0101rst\u0113\u0161anas rezult\u0101tus. Iev\u0113rojot \u0161\u012bs stingr\u0101s test\u0113\u0161anas proced\u016bras, ra\u017eot\u0101ji var nodro\u0161in\u0101t, ka vi\u0146u izstr\u0101d\u0101jumi ir uzticami, iztur\u012bgi un dro\u0161i lieto\u0161anai cilv\u0113kiem.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">K\u0101 \u201eCell Instruments\u201c var pal\u012bdz\u0113t<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Ra\u017eot\u0101jiem, kuri v\u0113las veikt prec\u012bzus <strong>st\u012bvums<\/strong> un <strong>iztur\u012bbas pret l\u016bzumiem testi<\/strong>, <strong>Cell Instruments Co., Ltd.<\/strong> pied\u0101v\u0101 pla\u0161u augstas precizit\u0101tes test\u0113\u0161anas iek\u0101rtu kl\u0101stu, tostarp <strong>Cauru\u013cu stingr\u012bbas testeri<\/strong> kas atbilst <strong>ISO 9626<\/strong> standarti. \u0160\u012bs ier\u012bces \u013cauj prec\u012bzi izm\u0113r\u012bt <strong>st\u012bvums<\/strong> un <strong>iztur\u012bba pret l\u016b\u0161anu<\/strong>, pal\u012bdzot ra\u017eot\u0101jiem izpild\u012bt kvalit\u0101tes kontroles pras\u012bbas un nodro\u0161in\u0101t, ka to produkti atbilst visaugst\u0101kajiem standartiem.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">M\u016bsu <strong>Cauru\u013cu stingr\u012bbas testeri<\/strong> ir izstr\u0101d\u0101ti, lai nodro\u0161in\u0101tu prec\u012bzus un atk\u0101rtojamus rezult\u0101tus, garant\u0113jot, ka j\u016bsu adatu caurules atbilst medic\u012bnisko ier\u012b\u010du meh\u0101niskaj\u0101m veiktsp\u0113jas pras\u012bb\u0101m. Izmantojot m\u016bsu test\u0113\u0161anas iek\u0101rtas, ra\u017eot\u0101ji var \u0101tri identific\u0113t jebk\u0101das probl\u0113mas saist\u012bb\u0101 ar cauru\u013cu veiktsp\u0113ju, t\u0101d\u0113j\u0101di nodro\u0161inot dro\u0161\u0101kus un efekt\u012bv\u0101kus medic\u012bniskos izstr\u0101d\u0101jumus.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Secin\u0101jums<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Port\u0101ls <strong>Adatas iztur\u012bbas tests<\/strong> ir b\u016btisks solis medic\u012bniskaj\u0101s ier\u012bc\u0113s izmantoto adatu cauru\u013cu darb\u012bbas r\u0101d\u012bt\u0101ju un dro\u0161\u012bbas nov\u0113rt\u0113\u0161an\u0101. Veicot t\u0101dus testus k\u0101, piem\u0113ram, <strong>cauru\u013cu st\u012bvuma tests<\/strong> un <strong>cauru\u013cu iztur\u012bba pret p\u0101rr\u0101vuma testu<\/strong>, ra\u017eot\u0101ji var nodro\u0161in\u0101t, ka to izstr\u0101d\u0101jumi atbilst visaugst\u0101kajiem dro\u0161\u012bbas un uzticam\u012bbas standartiem. Iev\u0113rojot <strong>ISO 9626<\/strong> nodro\u0161ina, ka caurule sp\u0113j iztur\u0113t meh\u0101nisk\u0101s slodzes, ar kur\u0101m t\u0101 saskarsies lieto\u0161anas laik\u0101, t\u0101d\u0113j\u0101di sniedzot medic\u012bnas speci\u0101listiem p\u0101rliec\u012bbu par izmantoto ier\u012b\u010du kvalit\u0101ti.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Lai nodro\u0161in\u0101tu uzticamu un prec\u012bzu test\u0113\u0161anu, <strong>Cell Instruments Co., Ltd.<\/strong> pied\u0101v\u0101 nepiecie\u0161amos r\u012bkus, lai izpild\u012btu \u0161os standartus un nodro\u0161in\u0101tu, ka j\u016bsu produkti medic\u012bniskaj\u0101 lieto\u0161an\u0101 darbojas optim\u0101li.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>","protected":false},"excerpt":{"rendered":"<p>Ievads adatu iztur\u012bbas test\u0101 Adatu iztur\u012bbas tests ir b\u016btisks, lai nov\u0113rt\u0113tu adatu cauru\u013cu meh\u0101nisk\u0101s \u012bpa\u0161\u012bbas, kuras tiek izmantotas injekciju adatu un citu medic\u012bnisko ier\u012b\u010du ra\u017eo\u0161an\u0101. Viens no \u0161\u012b testa galvenajiem aspektiem ir cauru\u013cu iztur\u012bbas pret l\u016b\u0161anu p\u0101rbaude, kas ir \u013coti svar\u012bga, lai nodro\u0161in\u0101tu dro\u0161\u012bbu un veiktsp\u0113ju [\u2026]<\/p>","protected":false},"author":5,"featured_media":0,"comment_status":"closed","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_uag_custom_page_level_css":"","site-sidebar-layout":"default","site-content-layout":"","ast-site-content-layout":"default","site-content-style":"default","site-sidebar-style":"default","ast-global-header-display":"","ast-banner-title-visibility":"","ast-main-header-display":"","ast-hfb-above-header-display":"","ast-hfb-below-header-display":"","ast-hfb-mobile-header-display":"","site-post-title":"","ast-breadcrumbs-content":"","ast-featured-img":"","footer-sml-layout":"","theme-transparent-header-meta":"","adv-header-id-meta":"","stick-header-meta":"","header-above-stick-meta":"","header-main-stick-meta":"","header-below-stick-meta":"","astra-migrate-meta-layouts":"set","ast-page-background-enabled":"default","ast-page-background-meta":{"desktop":{"background-color":"var(--ast-global-color-4)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""},"tablet":{"background-color":"","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""},"mobile":{"background-color":"","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""}},"ast-content-background-meta":{"desktop":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""},"tablet":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""},"mobile":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-gradient":""}},"footnotes":""},"categories":[26],"tags":[],"class_list":["post-2570","post","type-post","status-publish","format-standard","hentry","category-blog"],"acf":[],"uagb_featured_image_src":{"full":false,"thumbnail":false,"medium":false,"medium_large":false,"large":false,"1536x1536":false,"2048x2048":false,"trp-custom-language-flag":false},"uagb_author_info":{"display_name":"Editor","author_link":"https:\/\/www.pharmacopoeiatest.com\/lv\/author\/editor\/"},"uagb_comment_info":0,"uagb_excerpt":"Introduction to Needle Toughness Test The Needle Toughness Test is essential for evaluating the mechanical properties of needle tubing used in the manufacture of hypodermic needles and other medical devices. One of the key aspects of this test involves examining the resistance of tubing to breakage, which is critical for ensuring the safety and performance&hellip;","_links":{"self":[{"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/posts\/2570","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/users\/5"}],"replies":[{"embeddable":true,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/comments?post=2570"}],"version-history":[{"count":3,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/posts\/2570\/revisions"}],"predecessor-version":[{"id":2574,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/posts\/2570\/revisions\/2574"}],"wp:attachment":[{"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/media?parent=2570"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/categories?post=2570"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.pharmacopoeiatest.com\/lv\/wp-json\/wp\/v2\/tags?post=2570"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}